LeadCare Testing Systems (with Blood Obtained from a Vein) by Magellan Diagnostics: FDA Safety Communication - Risk of Inaccurate Results
- Discontinue using Magellan’s’ LeadCare System Testing Systems with venous blood samples. At this time, all LeadCare systems can be used with capillary blood samples.
- If laboratories or health care professionals are concerned about using the LeadCare Test Systems, the alternative options are mass spectrometry or atomic absorption methods. These are not point-of-care tests, and may be available only from larger-capacity laboratories such as reference labs.
- This Safety Communication replaces all previous communication from Magellan Diagnostics on their LeadCare lead testing systems including Magellan’s most recent Field Safety Correction Notification dated April 28, 2017.